REMOTE CONTROL
91
:SYSTem:CONFigure:OUTPut:PON[:STATe]
Set
Query
Description
Sets the output state at power-on. This is the
equivalent to the F-92 (Output Status when Power
ON) power on configuration settings.
Note
The setting will only be valid after the power has been
cycled.
Syntax
Return Syntax
:SYSTem:CONFigure:OUTPut:PON[:STATe]
{<NR1>|{SAFE|OFF}|{FORCe|ON}|AUTO}
:SYSTem:CONFigure:OUTPut:PON[:STATe]?
Parameter
SAFE | OFF | 0
The T3PS turns on in the same
state the unit was in prior to the
previous shut down. The output is
set to off (default).
FORCe | ON | 1
The T3PS turns on in the same
state the unit was in prior to the
previous shut down. The output is
set to on.
AUTO | 2
The T3PS turns on in the same
state the unit was in prior to the
previous shut down, but with the
same output on/off setting.
Return parameter 0
The power on output setting is
“SAFE” or “OFF”.
1
The power on output setting is
“FORCe” or “ON”.
2
The power on output setting is
“AUTO”.
Example
SYST:CONF:OUTP:PON SAFE
SYST:REB
Set the power-on output state to safe and reboot
the unit to active the setting.