REMOTE CONTROL
95
:SYSTem:CONFigure:CONTrol:RANGe
Set
Query
Description
This command is used to select the external analog
control voltage (or resistance) range. This is the
equivalent to the F-97 (Control Range) power on
configuration settings.
Note
The setting will only be valid after the power has been
cycled.
Syntax
Return Syntax
:SYSTem:CONFigure:CONTrol:RANG
{<NR1>|LOW|HIGH}
:SYSTem:CONFigure:CONTrol:RANGe?
Parameter
LOW | 0
5V [5kΩ]
HIGH| 1
10V [10kΩ]
Return Parameter <NR1>
Returns the range setting.
Example
SYST:CONF:CONT:RANG LOW
SYST:REB
Set external analog control voltage (or resistance)
range to 0~5V [5kΩ] and reboot the unit to active
the setting.
:SYSTem:CONFigure:TRIGger:INPut
:SOURce
Set
Query
Description
Sets or queries what action will be performed on
receiving a trigger.
Syntax
Return Syntax
:SYSTem:CONFigure:TRIGger:INPut:SOURce
{<NR1>|NONE|OUTPut|SETTing|MEMory}
:SYSTem:CONFigure:TRIGger:INPut:SOURce?
Parameter
NONE | 0
No input trigger.
OUTPut | 1
Toggles the output on receiving a
trigger.
SETTing | 2
Sets the voltage/current on receiving
a trigger.