4. Communication With a PC
74
Commands for setting the correction factor
Communication Command
Function
FR
Reads the number of the correction factor currently being
used.
Fn
Writes the number of the correction factor to be used.
n: Number of the correction factor to be used (0 to 15)
RFn
Reads the correction factor.
n: Number of the correction factor to be read (1 to 15)
Wn
_
xxxx_yyyy_zzzz_cccc
_: space
Writes the correction factor.
n: Number of the correction factor to be written (1 to 15)
xxxx: Correction factor X
2
yyyy: Correction factor Y
zzzz: Correction factor Z
cccc: Comment. Maximum 50 bytes.
The character of SPACE is not allowed.
CFn
Clears the correction factor.
n: Number of the correction factor to be cleared (1 to 15)
FKn
Switches between normal correction/direct correction.
n: 1 or 2, 1: Normal correction, 2: Direct correction
FKR
Reads the current correction type.
Return value 1: Normal correction 2: Direct correction
(direct
correction)
FAGn
Enables area correction.
n: Number of the area correction group (1 to 10)
FO (O: alphabets)
Disables area correction.
CGLn
Clears the chromaticity area in the area correction factor per
each individual group.
n: Number of the area correction group (1 to 10)
FGR
Reads the number of the area correction group currently
being used.
RGmKn
Reads the area correction factor.
m: Group Number (1 to 10) n: Area Number (1 to 5)
RGmLn
Reads the chromaticity area in the area correction factor.
m: Group Number (1 to 10) n: Area Number (1 to 5)
WGmKn
_
####
_
####
_
####
_: space
Writes the area correction factor.
m: Group Number (1 to 10) n: Area Number (1 to 5)
WGmLn
_
####
_
####
_
####
_: space
Writes the chromaticity area in the area correction factor.
m: Group Number (1 to 10) n: Area Number (1 to 5)
"
_
" indicates a space. "####" indicates a number.
When a communication command is sent from the PC, the instrument returns "OK" as the
reception acknowledgement command. When the instrument receives an illegal command, it
returns "NO".