5-39
Programming Commands
CALCulate2 Subsystem
WLIMit:STARt:WNUMber
Sets the starting wavenumber for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STARt:WNUMber {?|{<real> | MINimum | MAXimum}}
<
real
> is a wavenumber value that is within the following limits:
Constant
Description
MINimum 6060
cm
-1
MAXimum
wavelength limit stop value
Attribute
Summary
Non-sequential command
Preset
State: 6.060606E5 m
-1
*RST State: 6.060606E5 m
-1
SCPI Compliance: instrument specific
Description
This command sets the starting range for the wavelength limit. The default
units for the <
real
> parameter value are m
-1
. The start wavenumber value
must be less than or equal to the stop wavenumber value or the start wave-
number will be clipped to the stop wavenumber and a “Data out of range”
error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.