5-40
Programming Commands
CALCulate2 Subsystem
WLIMit:STOP:FREQuency
Sets the stopping frequency for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP:FREQuency {?|{<real> | MINimum | MAXimum }}
<
real
> is a frequency value that is within the following limits:
Constant
Description
MINimum start
wavelength
limit
MAXimum 428.2750
THz
Attribute
Summary
Non-sequential command
Preset
State: 249.8271 THz
*RST State: 249.8271 THz
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The default
units for the <
real
> parameter are Hz. The stop frequency value must be
greater than or equal to the start frequency value or the stop frequency will be
clipped to the start frequency and a “Data out of range” error will be gener-
ated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.