Section 7. Installation
232
7.7.12.4.2
Two-Point Calibrations (gain and offset)
Use this two-point calibration procedure to adjust multipliers (slopes) and offsets
(y intercepts). See FieldCal() Slope and Offset (Opt 2) Example
(p. 238)
and
FieldCal() Slope (Opt 3) Example
(p. 240)
for demonstration programs:
1. Use a separate FieldCal() instruction and separate variables for each sensor to
be calibrated.
2.
Ensure mode variable =
0 or 6 before starting.
a. If Mode > 0
and ≠
6, calibration is in progress.
b. If Mode < 0, calibration encountered an error.
3. Place sensor into first known point condition.
4. Set KnownVar variable to first known point.
5. Set Mode
variable =
1 to start first part of calibration.
a. Mode
=
2 (automatic) during the first point calibration.
b. Mode
=
3 (automatic) when the first point is completed.
6. Place sensor into second known point condition.
7. Set KnownVar variable to second known point.
8. Set Mode
=
4 to start second part of calibration.
a. Mode
=
5 (automatic) during second point calibration.
b. Mode
=
6 (automatic) when calibration is complete.
7.7.12.4.3
Zero Basis Point Calibration
Zero-basis calibration (FieldCal() instruction Option 4) is designed for use with
static vibrating wire measurements. It loads values into zero-point variables
to track conditions at the time of the zero calibration. See FieldCal() Zero
Basis (Opt 4) Example
(p. 243)
for a demonstration program.
7.7.12.5 Field Calibration Examples
FieldCal() has the following calibration options:
•
Zero
•
Offset
•
Two-point slope and offset
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......