Section 7. Installation
280
CRBasic Programs and Notes
PT100 BrHalf4W() Four-Wire Half-Bridge Calibration
'This program example demonstrates the calibration of a 100-ohm PRT (PT100) in a four-wire
'half bridge with voltage excitation. See adjacent procedure and schematic.
'Declare constants and variables:
Const
Rf = 100000
'Value of bridge resistor
Public
X
'Raw output from the bridge
Public
RS0
'Calculated PT100 resistance at 0 °C
BeginProg
Scan
(1,Sec,0,0)
....'Measure X:
'BrHalf4W(Dest,Reps,Range1,Range2,DiffChan,ExChan,MeasPEx,ExmV,RevEx,RevDiff,
' SettlingTime,Integ,Mult,Offset)
BrHalf4W
(X,1,mV50,mV50,1,Vx1,1,4160,True,True,0,250,1,0)
'Calculate RS0:
RS0 = X * Rf
NextScan
EndProg
PT100 BrHalf4W() Four-Wire Half-Bridge Measurement
'This program example demonstrates the measurement of a 100-ohm PRT in a four-wire
'half bridge using current excitation. See previous procedure and schematic.
'Declare constants and variables:
Const
Rf = 100000
'Value of bridge resistor
Const
RS0 = 100000
'Resistance of PT100 at 0 °C from calibration program
Public
X
'Raw output from the bridge
Public
RS
'Calculated PT100 resistance
Public
RS_RS0
'Calculated ratio of RS/RS0
Public
DegC
'Calculated temperature
BeginProg
Scan
(1,Sec,0,0)
....'Measure X:
'BrHalf4W(Dest,Reps,Range1,Range2,DiffChan,ExChan,MeasPEx,ExmV,RevEx,RevDiff,
' SettlingTime,Integ,Mult,Offset)
BrHalf4W
(X,1,mV50,mV50,1,Vx1,1,4160,True,True,0,250,1,0)
'Calculate RS and RS/RS0:
RS = X * Rf
RS_RS0 = RS/RS0
....'Calculate temperature from RS_RS0:
'PRTCalc(Dest,Reps,Source,PRTType,Mult,Offset)
PRTCalc
(DegC,1,RS_RS0,1,1.0,0)
NextScan
EndProg
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......