PMX command set (API)
434
A4354-2.2
HBM: public
PMX
Parameters:
p1: Peak-value memory 1 (Max), 2 (Min) or 3 (Peak-
toPeak)
p2=1: suspend update
p2=0: enable update (default)
Every time you switch on, the status is set to "enable ap
dating".
Virtual subchannels (slot9 / channel 9), digital subchan
nels (slot10 / channel 10) and the PX878 multi IOcard
don’t support peak values.
Note:
The response depends on the SRB command.
Hold Peak Value Query
Read out the peak-value memory update status of all
selected channels (PCS/SPS)
Syntax1:
HPV?p1(x)
Parameters:
p1: Peak-value memory 1 (Max), 2 (Min) or 3 (Peak-
toPeak)
Response:
q1,q2(y):
q1: Requested peak-value memory
q2: 1: Update interrupted
0: Update activated: e. g. for Max (p1=1) 1,1
:
1,0
::
1,1
::
indicates that this subchannel (between the two
:
) has
no max, min or peak-topeak values
Syntax 2:
HPV??(x)
(y): Available peak-value memories: e. g. 1,2,3
:
1,2
::
1
::
indicates that this subchannel (between the two
:
) has no
peak values
The command returns the status of the peak-value
memory, which can be set by the HPV command.
HPV?
Summary of Contents for PX401
Page 1: ...Operating Manual English PMX...
Page 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Page 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Page 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Page 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Page 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Page 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Page 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Page 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Page 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Page 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Page 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Page 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Page 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Page 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Page 489: ...Object directory PMX A4354 2 2 HBM public 489...
Page 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...