CONFigure
The
CONFigure
command subsystem configures the multimeter to perform the
specified measurement with the given range and resolution.
CONFigure
does not
make the measurement after setting the configuration. Use the
INITiate
command to
place the multimeter in the wait-for-trigger state and store readings in mainframe or
command module memory. Or, use the
READ?
command to make the measurement
and send the readings to the output buffer when the trigger is received.
Executing
CONFigure
is equivalent to configuring the multimeter with the low-level
commands shown in the following table.
Command
Setting
VOLTage:RANGe
RESistance:RANGe
As specified, or autorange.
VOLTage:RESolution
RESistance:RESolution
As specified, or as a function of range, integration time,
or aperture time.
VOLTage:APERture
RESistance:APERture
16.7 ms (60 Hz) or 20 ms (50 Hz), or based on specified
resolution.
VOLTage:NPLC
RESistance:NPLC
1 PLC, or based on specified resolution.
CALibration:ZERO:AUTO
ON (autozero is performed after every measurement).
RESistance:OCOMpensated
OFF (applies to resistance measurements only).
TRIGger:SOURce
TRIGger:COUNt
TRIGger:DELay
IMM (trigger signal is always true).
1
AUTO (DC volts/resistance: 0 s, AC volts: 0.5 s).
SAMPle:COUNt
SAMPle:SOURce
1
IMM
Subsystem Syntax
CONFigure
:FRESistance [<range>[,<resolution>]] [,<channel_list>]
:RESistance [<range>[,<resolution>]] ,<channel_list>
:TEMPerature <transducer>,<type> ,<channel_list>
:VOLTage:AC [<range> [,<resolution>]] [,<channel_list>]
:VOLTage[:DC] [<range> [,<resolution>]] [,<channel_list>]
NOTE: If range and resolution are not specified (that is, if you use the default
values), use a space rather than a comma before the channel_list parameter.
Chapter 5
Command Reference 126