Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
<
source
>
discrete
BUS | EXT | HOLD | IMMediate |
TTLTrg0 - TTLTrg7
none
Comments
•
The
TRIGger:SOURce
command only selects the trigger source. You must use
the
INITiate
command to place the multimeter in the wait-for-trigger state. (The
MEASure
command automatically executes an
INITiate
command.)
•
TRIGger:SOURce EXT
uses the multimeter’s External Trigger BNC connector as
the trigger source. The multimeter triggers on the falling (negative-going) edge of
a TTL input signal.
•
TRIGger:IMMediate
causes a trigger to occur immediately provided the
multimeter is placed in the wait-for-trigger state using
INITiate
.
•
When
TRIGger:SOURce BUS
is selected,
ABORt
returns the multimeter to the
idle state. When a Group Execute Trigger (
GET
) bus command or
*TRG
common
command is executed, the “Trigger ignored” error is generated.
•
When
TRIGger:SOURce HOLD
is selected,
ABORt
returns the multimeter to the
idle state. All subsequent single triggers sent using
TRIGger:IMMediate
are
ignored and the “Trigger ignored” error is generated.
•
The
CONFigure
and
MEASure
command subsystems automatically set the trigger
source to
TRIG:SOUR IMM
.
•
The
READ?
command cannot be used if the trigger source is
TRIG:SOUR BUS
or
TRIG:SOUR HOLD
.
•
Related Commands:
ABORt
,
INITiate
,
TRIGger:IMMediate
•
*RST Condition:
TRIG:SOUR IMM
Example
Setting the Sample Source
CONF:VOLT:DC (@100:104)
!Function: DC voltage;
specify channel list.
TRIG:SOUR EXT
!Trigger source is external BNC on
multimeter front panel.
TRIG:COUN 10
!Multimeter will accept 10 external triggers
(5 channels will be scanned with each
trigger).
READ?
!Place multimeter in wait-for-trigger state;
make measurements when external trigger is
received; send readings to output buffer.
enter statement
!Enter readings into computer.
:SOURce?
TRIGger:SOURce?
returns “
BUS
”, “
EXT
”, “
HOLD
”, or “
IMM
” to show the present
trigger source. The quoted string is sent to the output buffer.