24 Integra Series
Quick Start Guide
2-Wire Ohms
SENS:RES:AVER:WIND <NRf>
‘ Set filter window in %: <NRf> = 0 to 10.
SENS:RES:AVER:COUN <n>
‘ Set filter count: <n> = 1 to 100.
SENS:RES:AVER:TCON <name>
‘ Select filter type: <name> = MOV or REP.
SENS:RES:AVER:STAT <b>
‘ Enable or disable filter: <b> = ON or OFF.
4-Wire Ohms
SENS:FRES:AVER:WIND <NRf>
‘ Set filter window in %: <NRf> = 0 to 10.
SENS:FRES:AVER:COUN <n>
‘ Set filter count: <n> = 1 to 100.
SENS:FRES:AVER:TCON <name>
‘ Select filter type: <name> = MOV or REP.
SENS:FRES:AVER:STAT <b>
‘ Enable or disable filter: <b> = ON or OFF.
Temperature
SENS:TEMP:WIND <NRf>
‘ Set filter window in %: <NRf> = 0 to 10.
SENS:TEMP:AVER:COUN <n>
‘ Set filter count: <n> = 1 to 100.
SENS:TEMP:AVER:TCON <name>
‘ Select filter type: <name> = MOV or REP.
SENS:TEMP:AVER:STAT <b>
‘ Enable or disable filter: <b> = ON or OFF.
Remote programming example (filter for 2-wire ohms)
SENS:RES:AVER:WIND 2
‘ Sets filter window to 2%.
SENS:RES:AVER:COUN 10
‘ Set filter count to 10.
SENS:RES:AVER:TCON REP
‘ Select Repeat filter type.
SENS:RES:AVER:STAT ON
‘ Enable filter.
NOTE
When finished with remote programming, press the LOCAL key to
restore front panel control.
Line cycle synchronization (lsync)
Synchronizing A/D conversions with the frequency of the power line increases
common mode and normal mode noise rejection. When line cycle synchronization
is enabled, the measurement is initialized at the first positive-going zero crossing
of the power line cycle after the trigger.
Enabling or disabling line cycle synchronization
1.
Referring to
, press SHIFT and then RATE (LSYNC) to display the
present state of line synchronization (off or on).
2.
Use the edit keys to (see
) enable (ON) or disable (OFF) line syn-
chroniztion.
SCPI programming commands (lsync)
SYST:LSYN:STAT <b>
‘ Enable or disable line synchronization: <b> = ON or OFF.
Remote programming example (lsync)
SYST:LSYN:STAT ON
‘ Enable line synchronization.
NOTE
When finished with remote programming, press the LOCAL key to
restore front panel control.
To
p
To
p
To
p
Summary of Contents for Integra 27 Series
Page 3: ......
Page 8: ...QuickStartGuide Integra Series...