10-12
SCPI Command Reference
Models 2510 and 2510-AT User’s Manual
Table 10-7
:STATus subsystem commands
Command
Description
Default
parameter
SCPI
:STATus
Note 1
✓
:MEASurement
Path to control measurement event registers.
[:EVENt]?
Read the event register.
6
Note 2
✓
:ENABle <NRf> | <NDN>
Program the enable register.
Note 3
✓
:ENABle?
Read the enable register.
6
✓
:CONDition?
Read the condition register.
6
✓
:OPERation
Path to control operation event registers.
✓
[:EVENt]?
Read the event register.
6
Note 2
✓
:ENABle <NRf> | <NDN>
Program the enable register.
Note 3
✓
:ENABle?
Read the enable register.
6
✓
:CONDition?
Read the condition register.
6
✓
:QUEStionable
Path to control questionable event registers.
✓
[:EVENt]?
Read the event register.
6
Note 2
✓
:ENABle <NRf> | <NDN>
Program the enable register.
Note 3
✓
:ENABle?
Read the enable register.
6
✓
:CONDition?
Read the condition register.
6
✓
:PRESet
Return status registers to default states.
✓
:QUEue
Path to access error queue.
[:NEXT]?
Read the most recent error message.
Note 4
✓
:ENABle <list>
Specify error and status messages for queue.
Note 5
✓
:ENABle?
Read the enabled messages.
✓
:DISable <list>
Specify messages not to be placed in queue.
Note 5
:DISable?
Read the disabled messages.
:CLEar
Clears all messages from Error Queue.
Notes:
1. Commands in this subsystem are not affected by *RST and :SYSTem:PREset. The effects of cycling power, *CLS and
:STATus:PRESet, are explained by the following notes.
2. Event Registers — Power-up and *CLS clears all bits.:STATus:PRESet has no effect.
3. Enable Registers — Power-up and :STATus:PRESet clears all bits. *CLS has no effect. Accepts the SCPI 1995.0 mandated (non-
decimal numeric) format (#H, #Q, or #B).
4. Error Queue — Power-up and *CLS clears all bits of the registers.
5. Error Queue Messages — Power-up clears list of messages. *CLS and :STATus:PRESet have no effect.
6. Register Query Commands — The format for the response messages (ASCII, hexadecimal, octal, or binary) depends on which
data format is presently selected. See the :FORMat:SREGister command.
Summary of Contents for TEC SourceMeter 2510
Page 194: ...A Specifications...
Page 197: ...B StatusandErrorMessages...
Page 206: ...C IEEE 488BusOverview...
Page 221: ...D IEEE 488andSCPI ConformanceInformation...
Page 225: ...E ExamplePrograms...
Page 233: ...F GPIB488 1Protocol...