10-38
SCPI Command Reference
Models 2510 and 2510-AT User’s Manual
DEFault
50
Ω
(100
Ω
RTD or thermistor)
5e2
Ω
(1k
Ω
RTD or thermistor)
5e3
Ω
(10k
Ω
thermistor)
5e4
Ω
(100k
Ω
thermistor)
:STATe <b> = ON | OFF
Enable/disable protection
ON | 1
Enable protection
OFF | 0
Disable protection
Query
[:HIGH][:LEVel]?
Query high limit
[:HIGH]:TRIPped?
Query if high limit exceeded (1 = exceeded)
:LOW[:LEVel]?
Query low limit
:LOW:TRIPped?
Query if low limit exceeded (1 = exceeded)
:STATe?
Query if protection is enabled.
Description
The :HIGH and :LOW commands program the high and low limits
respectively for the resistance function, while the :STATe command
enables and disables resistance function limits. The two :TRIPPed? que-
ries return a value of 1 if the corresponding limit is exceeded or a value
of 0 if it is not exceeded. :STATe? returns a value of 1 if protection is
enabled. Note that the high limit cannot be lower than the low limit, and
the low limit cannot be higher than the high limit.
Temperature function
[:SPOint] <n>
:SOURce[1]:TEMPerature:[SPOint] <n>
Program temperature setpoint
Parameters
<n> =
-50 to +225°C
Set temperature setpoint
223.15 to 498.15K
-58 to 437°F
MINimum
-50°C
223.15K
-58°F
MAXimum
+225°C
498.15K
437°F
DEFault
+25°C
298.15K
77°F
Query
[:SPOint]?
Query temperature setpoint
Description
This command programs the temperature setpoint for the temperature
function. This value is the temperature at which the Model 2510 will
maintain system temperature when the temperature function is enabled.
The setpoint cannot exceed the temperature protection limits.
Summary of Contents for TEC SourceMeter 2510
Page 194: ...A Specifications...
Page 197: ...B StatusandErrorMessages...
Page 206: ...C IEEE 488BusOverview...
Page 221: ...D IEEE 488andSCPI ConformanceInformation...
Page 225: ...E ExamplePrograms...
Page 233: ...F GPIB488 1Protocol...