Models 2510 and 2510-AT User’s Manual
SCPI Command Reference
10-49
:POSetup
:SYSTem:POSetup <name>
Program power-on defaults
Parameters
<name> =
RST
Power-up to *RST defaults
PRESet
Power-up to :SYSTem:PRESet defaults
SAV0
Power-up to setup stored at memory
location 0
SAV1
Power-up to setup stored at memory
location 1
SAV2
Power-up to setup stored at memory
location 2
SAV3
Power-up to setup stored at memory
location 3.
SAV4
Power-up to setup stored at memory
location 4.
Query
:POSetup?
Query power-on setup
Description
This command is used to select the power-on defaults. With RST
selected, the instrument powers up to the *RST default conditions. With
PRES selected, the instrument powers up to the :SYStem:PRESet
default conditions. Default conditions are listed in the
through
.
With the SAV0-4 parameters specified, the instrument powers-on to the
setup that is saved in the specified location using the *SAV command.
Select power line frequency setting
:LFRequency <freq>
:SYSTem:LFRequency <freq>
Select line frequency
Parameters
<freq> =
50
50Hz setting
60
60Hz setting
Query
:LFRequency?
Query line frequency selection.
Description
Use this command to manually select the line frequency setting (50 or
60Hz). The line frequency setting should match the power line fre-
quency, or readings may be noisy.