Models 2510 and 2510-AT User’s Manual
SCPI Command Reference
10-21
Data elements
:ELEMents <item list>
:FORMat:ELEMents <item list>
Specify data elements for data string
Parameters
<item list> = VOLTage
Includes voltage reading
CURRent
Includes current reading
POWer
Includes power reading
RESistance
Includes resistance reading
TEMPerature
Includes temperature reading
TSENsor
Includes sensor reading
TIME
Includes timestamp
STATus
Includes status information
NOTE
Each item in the list must be separated by a comma (i.e., :ELEMents, VOLTage,
CURRent, RESistance).
Query
:ELEMents?
Query elements in data string
Description
This command is used to specify the elements to be included in the data
string in response to the following queries:
:FETCh?
:READ?
:MEASure[:<function>]?
You can specify from one to all five elements. Each element in the list
must be separated by a comma (,). These elements (shown in
) are explained as follows:
NOTE
An overflow reading reads as +9.9E37.
VOLTage
— This element provides the TEC voltage measurement.
CURRent
— This element provides the TEC current measurement.
RESistance
— This element provides the TEC resistance measurement.
TEMPerature
— This element provides the TEC temperature
measurement.
TSENSor
— This element provides the sensor temperature
measurement.
TIME
— A timestamp is available to reference each group of readings
to a point in time. The relative timestamp operates as a timer that starts
at zero seconds when the instrument is turned on or when the relative
timestamp is reset (:SYSTem:TIME:RESet). The timestamp for each
reading sent over the bus is referenced, in seconds, to the start time.
STATus
— Each and every measurement performed by the Model 2510
has a STATus word associated with it. This STATus word is accessible
via the :FORMat:ELEMents STATus element. Internally, the Model
2510 maintains this STATus element as a 16-bit integer. Each bit within
this integer provides status information about a particular process asso-
ciated with the construction of the measurement. However, when que-
ried, the decimal equivalent is returned. The format of the STATus
element is as follows: