M9506A-BC1 5-Slot AXIe Chassis with Integrated Controller User Guide
63
SCPI Programming Interface
4
:CONFigure:MEASurement:SETTing:ACTual?
Query
:CONFigure:MEASurement:SETTing:ACTual?
Description
This query returns the current measurement setting operating in digitizer,
std_int32 [1-10], -1 if the current setting is changed and not saved.
Example
-> conf:meas:sett:act?
<- +1
:CONFigure:MEASurement:SETTing:CANCel
Command
:CONFigure:MEASurement:SETTing:CANCel
Description
This query resets the present measurement config back to previously saved
measurement config loaded. To go back to default, a module reset is required if
the default setting is not saved in the slot.
Example
-> conf:meas:sett:canc
:DIGitizer[N]:TRIGger:DELay[?]
Command
:DIGitizer[N]:TRIGger:DELay
Description
This command sets acquisition delay after the trigger event occurs; only positive
delay values are accepted, std_float32.
Example
-> dig3:trig:del 0.0
Query
:DIGitizer[N]:TRIGger:DELay?
Description
This query returns the current trigger delay setting, std_float32.
Example
-> dig3:trig:del?
<- +0.00000
:DIGitizer[N]:TRIGger:INPut[?]
Command
:DIGitizer[N]:TRIGger:INPut
Description
This command selects one of several trigger options. The value entered as quoted
std_string.
- Immediate – Trigger event happens immediately upon ARM command.