1269 User Manual
SCPI Command Reference 5-7
Table 5-1, VXIbus Model 1269 EMS SCPI Commands List Summary
Keyword
Parameter Form (Default in Bold)
SCPI 1994.0 Notes
SYSTem
Confirmed
:ERRor ?
Confirmed
(Query only)
:VERSion?
Confirmed
(Query only)
:COMMunicate
Confirmed
:SERial
Confirmed
[:RECeive]
Confirmed
:BAUD
1200 | 2400 | 4800 |
9600
Confirmed
Combinations ofParity/Data/StopBits
:BITS
7 |
8
Confirmed
which are NOT allowed:
:SBITs
1
| 2
Confirmed
NONE/7/1
:PARity
Confirmed
EVEN/8/2
[:TYPE]
EVEN | ODD |
NONE
Confirmed
ODD/8/2
:TRANsmit
Confirmed
:AUTO
Confirmed
Programming either the TRANSMIT
:BAUD
1200 | 2400 | 4800 |
9600
Confirmed
or the RECEIVE direction programs
:BITS
7 |
8
Confirmed
BOTH directions
:SBITs
1
| 2
Confirmed
:TIME
<hour>, <minute>, <second>
Confirmed
ranges: 0 - 23, 0 - 59, 0 - 59
:DATE
<year>, <month>, <day>
Confirmed
ranges: 1995 - 2120, 1 - 12, 1 - 31
:MEASure
Confirmed
:VOLTage[<N>]?
Confirmed*
<N> = 1 - 7 (Query only)
:FREQuency[<N>]?
Confirmed*
<N> = 1 (Query only)
:TIME[<N>]?
Not Confirmed
<N> = 1 - 3 (Query only)
:TEMPerature[<N>}?
Confirmed*
<N> = 1 - 14 (or 1 - 27) (Query only)
:VXI
Not Confirmed
:ACFail?
Not Confirmed
(Query only)
:ASTRobe?
Not Confirmed
(Query only)
:SYSFail?
Not Confirmed
(Query only)
STATus
Confirmed
:PRESet
Confirmed
(Command only)
:OPERation
Confirmed
Command has no effect
[:EVENt]?
Confirmed
Command has no effect (Query only)
:CONDition?
Confirmed
Command has no effect (Query only)
:ENABle
(0;0;32767)
Confirmed
Command has no effect
:QUEStionable
Confirmed
[:EVENt]?
Confirmed
(Query only)
:CONDition?
Confirmed
(Query only)
:ENABle
(0;0;32767)
Confirmed
:VOLTage
Confirmed
[:EVENt]?
Confirmed
(Query only)
:CONDition?
Confirmed
(Query only)
:ENABle
(0;0;32767)
Confirmed
Meaningful range is 0 to 127
:TEMPerature
Confirmed
[:EVENt]?
Confirmed
(Query only)
:CONDition?
Confirmed
(Query only)
:ENABle
(0;0;32767)
Confirmed
Meaningful range is 0 to 16383
:FREQuency
Confirmed
[:EVENt]?
Confirmed
(Query only)
:CONDition?
Confirmed
(Query only)
:ENABle
(0;0;32767)
Confirmed
Meaningful range is 0 to 15
:TIME
Confirmed
[:EVENt]?
Confirmed
(Query only)
:CONDition?
Confirmed
(Query only)
:ENABle
(0;0;32767)
Confirmed
Meaningful range is 0 to 7
:VXI
Not Confirmed
[:EVENt]?
Not Confirmed
(Query only)
:CONDition?
Not Confirmed
(Query only)
:ENABle
(0;0;32767)
Not Confirmed
Meaningful range is 0 to 1024
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com