:measure:voltage:AC
Syntax:
:measure:voltage:AC
□
{<
range>
|MIN|MAX|DEF}
Function:
The command sets the measure range of AC voltage.
Parameters:
Name
Type
Range of Value
default
<
range>
discrete
{0|1|2|3|4}
2
Explanations:
The Table 2- 3 lists the parameters and the ranges.
Table 2- 3 The range of AC voltage
Parameter
Range
0
200mV
1
2.0V
2
20.0V
3
200.0V
4
750.0V
MIN
200mV
MAX
750V
DEF
20.0V
Example
:
To set the AC voltage range as 750.0V:
:measure:voltage:AC
□
MAX or :measure:voltage:AC
□
4
12.
:measure:voltage:AC:range?
Syntax:
:measure:voltage:AC:range?
Function:
The command queries the measure range of AC voltage.
Return Format: