Programming Guide for DM3000 Series
2-22
26.
:measure:current:AC:digit
Syntax:
:measure:current:AC:digit?
:measure:current:AC:digit
□
{INC|DEC|5|6|7}
Function:
The command queries and sets the digit of AC current.
Explanations:
:measure:current:AC:digit? queries the digit of the AC current. Before sending the
command, please turn on AC current measure function.
:measure:current:AC:digit
□
{INC|DEC|5|6|7} sets the digit of the AC current.
The Table 2- 2 indicates the parameters and their meanings.
Remarks:
If the showing digit reachs 5, DEC command will be invalid. If 7,INC
invalid.
Return Format:
:measure:current:AC:digit? returns the current digit: 5
、
6 or 7.
Example:
To set the digit as 7:
:measure:current:AC:digit
□
7
27.
:measure:current:AC:freq?
Syntax:
:measure:current:AC:freq?
Function:
The command queries the measure frequency of AC current.
Return Format:
The query returns the current frequency in the form of 5.01, and unit is Hz.
(for example)
28.
:measure:current:AC:freq:display
Syntax
;