RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-25
33.
:measure:resistance:range?
Syntax:
:measure:resistance:range?
Function:
The command queries the measure range of two circuits resistance.
Return Format:
The query returns the current range parameter: 0, 1, 2, 3, 4, 5 or 6.(see
Table 2- 6)
34.
:measure:resistance:digit
Syntax:
:measure:resistance:digit?
:measure:resistance:digit
□
{INC|DEC|5|6|7}
Function:
The command queries and sets the digit of two circuits resistance.
Explanations:
:measure:resistance:digit? queries the digit of the two circuits resistance.
:measure:resistance:digit
□
{INC|DEC|5|6|7} sets the digit of the two circuits
resistance.
Remarks:
If the showing digit reachs 5, DEC command will be invalid. If 7,INC
invalid.
Return Format:
:measure:resistance:digit? returns the digit of the current resistance: 5
、
6 or 7.
Example:
To set the digit as 6:
:measure:resistance:digit
□
6
35.
:measure:fresistance?
Syntax:
:measure:fresistance?