Programming Guide for DM3000 Series
2-77
1.
:datalog
Syntax:
:datalog?
Function:
The command queries whether the datalog function is on.
Explanation:
Befroe the datalog function is on, please configure the datalog settings.
Return Format:
The return value “Run” and “Stop” respectively denote that the datalog system runs or
stops.
2.
:datalog:configure:function
Syntax:
:datalog:configure:function?
:datalog:configure:function
□
<DCV,DCI,2WR,4WR>,<range>
Function:
The command queries and sets the datalog function.
Explanations:
:configure:function? queries the current datalog function.
:configure:function
□
<DCV,DCI,2WR,4WR>,< range>
sets the datalog function. The
first parameter including DCV,DCI,2WR and 4WR shows the datalog type.; the second
paramater shows the correspanding range.
Remarks:
the auto range is unavailable for the datalog function. The lower level of
the range is 1; and the upper level is related with the set function. The upper level of
DCV range is 5; the upper level of DCI is 5; the upper level of 2WR is 7; and the upper
level of 4WR is 7.
Return Format:
:datalog:configure:function? returns the corresponding range parameters of
DCV,DCI,2WR or 4WR.