ST1952 Operation Manual
50
:DATA?
Command syntax:
:CALCulate2:DATA?
Description:
This query command is used to read the result of the CALC2 calculation. If CALC2 is disabled or
NONE is selected, the “raw” reading will be read.
:CALCulate3
These commands are used to configure and control the CALC3 limit test.
:LIMit[1]:UPPer <n>
Specify upper limit
:LIMit[1]:LOWer <n>
Specify lower limit
Command syntax:
:CALCulate3:LIMit[1]:UPPer <n>
:CALCulate3:LIMit[1]:LOWer <n>
Command Parameter:
<n> = -100e6 to 100e6
Specify limit value
DEFault Specify upper limit to 1
Specify lower limit to -1
MINimum Specify limit to -100e6
MAXimum
Specify limit to +100e6
Query:
:UPPer?
Query upper limit
:LOWer?
Query lower limit
Description:
This command is used to specify the upper and lower limit for LIMIT1. The actual limit depends on
which measurement function is currently selected. For example, a limit value of 1 is 1V for the volts
functions (DCV or ACV), 1A for the current function (DCI or ACI), 1Ω on the ohms functions (2 or 4).
A limit value is not range sensitive. A limit of 1 for DCV is 1V on all measurement ranges.
:STATe <b>
Command syntax:
:CALCulate3:LIMit[1]:STATe <b>
Command Parameter:
<b> = 0 or OFF
Disable limit test
1 or ON
Enable limit test
Query:
:STATe?
Query state (on or off) of limit test