4
IM 701430-51E
Activating the Trigger on Fields and Frames
Specify the type of CAN signal field or frame to be used as trigger conditions from the
following five types. The trigger is activated on the AND condition of the five types
(combination trigger). However, RTR and Data Field cannot be specified in combination.
• Start of Frame: Activates the trigger on the Start of Frame. The trigger point is set
to the end of the Start of Frame.
• Identifier:
Activates the trigger on an Identifier matching the specified
condition. The trigger point is set to the end of the Identifier.
• RTR:
Activates the trigger on a remote frame (RTR is recessive). The
trigger point is set to the end of the RTR bit.
• Data Field:
Activates the trigger on a data field matching the specified
condition. The trigger point is set to the end of the data field.
• Error Frame:
Activates the trigger on an error frame. The DL7200 considers 6
successive dominant bits as an error frame trigger condition.
Therefore, a trigger occurs if 6 successive dominant bits occur in
an overload frame. The trigger point is set to the end of the 6
th
bit.
If multiple field and frame types are combined, the trigger point is set to the trigger point
of the type that appeared last in the time sequence.
If you specified Identifier or Data Field, specify the bit pattern.
When using hexadecimal notation, enter X or 0 through F in units of 4 bits.
When using binary notation, enter X, 0, or 1.
The following figure shows the field and frame formats and trigger position.
S: Trigger position of the Start of Frame
I: Trigger position of the Identifier
R: Trigger position of RTR
D: Trigger position of the Data Frame
E: Trigger position of the Error Frame
• Data frame
Interframe
Space
Interframe
Space or
Overload
Frame
Recessive
Dominant
DATA FRAME
Start of Frame
(1 bit)
S
D
Arbitration Field
(Std: 12 bit)
(Ext: 32 bit)
Control Field
(6 bit)
Data Field
(0 to 64 bit)
CRC Field
(CRC Sequence 15 bit)
(CRC Delimiter 1 bit)
ACK Field
(ACK Slot 1 bit)
(ACK Delimiter 1 bit)
End of Frame
(7 bit)
2. Setting the Trigger Conditions