7351 Series Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-32
NOTE:
When using the SCPI commands via USB, note the following point.
After sending a setting command, send a query command and receive its response.
Sending more than one setting command in succession may cause unintended behavior or an error to
occur.
As query command for setting commands, error check by ":SYSTem:ERRor?" and completion check by
"*OPC?" are generally used.
1. Measurement function and trigger system relevant command
Item
SCPI command
Description
Initial
value
Measurement Measurement
function
[:SENSe]:FUNCtion
“<function>”
“<function>”:
“VOLTage:DC”
DC voltage measurement (DCV)
“VOLTage:AC”
AC voltage measurement (ACV)
“RESistance”
Resistance measurement (2W
)
“CURRent:DC”
DC current measurement (DCI)
“CURRent:AC”
AC current measurement (ACI)
“VOLTage:ACDC”
AC voltage (AC+DC coupling) mea-
surement (ACV(AC+DC))
“CURRent:ACDC”
AC current (AC+DC coupling)
measurement (ACI(AC+DC))
“DIODe”
Diode measurement (Diode)
“RESistance:LPOWer”
Low power resistance measure-
ment (LP-2W
“CONTinuity”
Continuity test (Cont)
“FREQuency”
Frequency measurement (Freq)
[:SENSe]:FUNCtion? Reply: “VOLT:DC”, “VOLT:AC”, “RES”,
“CURR:DC”,CURR:AC”, “VOLT:ACDC”,
“CURR:ACDC”, “DIOD”,”RES:LPOW”, “CONT” or
“FREQ”
:CONFigure:<function
{<real>|MAXimum
|MINimum} *1
Changing to the specified measurement function
<real>:
Range setting (0
Setting value
Maximum value for the
measurement range)
(For the frequency measurement, the maximum value is
750(V))
This function cannot be set for the continuity and diode mea-
surements.
:CONFigure?
Reply: "Fud.ddddddE
dd"
+d.ddddddE
dd" shows a measurement range.
*1: <Function>: Specifies a function, which is not enclosed by double quotes, for "measurement function".