25 - 40
021774/07/14
6.3 Commands and description
The following table lists the available commands and the associated passwords for read / write:
Command
Initial value
factory
setting
MODBUS
register-
address
Description
Password
Read
1
/ Write
2
Command AI
10
40069
Averaging interval for wind speed and
wind direction.
None
User
Command BR
96
40005
Select baud rate.
None
User
Command CI
0
40013
Command interpreter.
None
User
Command DC
0
40081
Mode of calculation for twilight.
None
User
Command DO
0
40037
North correction of wind direction.
None
User
Command FB
1
40001
Quick-start mode.
None
User
Command ID
0 (THIES)
1 (MODBUS)
40003
Identification number resp. slave
address.
None
User
Command KY
0
40009
Set key / password.
None
None
Command LC
0
40045
LED control.
None
None
Command RS
-
40029
Reset.
None
User
Command SH
0
40071
Station height.
None
User
Command SV
-
45005
SW-Version.
None
-
Command TR
-
-
Telegram output.
None
None
Command TT
0
-
Automatic telegram output.
None
None
Command TZ
0
40073
Time zone.
None
User
Table 7 : List of commands
1
: Command without parameter (used to read selected parameter).
2
: Command with parameter (used to write a new parameter).
6.3.1 Command AI
<id>AI<parameter><CR>
Averaging interval for wind speed and wind direction
Access:
Read / write
Description:
The command AI is used to specify the averaging interval for the wind
speed and wind direction in minutes.
The averaging of the wind velocity is scalar, and the averaging of the
wind direction vectorial.
If the parameter is 0, the averaging is deactivated, and the mean values
correspond to the instantaneous values. Here, the wind direction is reset
(to 0°) during calm (< 0.6m/s).
Wind from the North is displayed with 360°.
Parameter description:
AI = 0
averaging disabled
AI = 1
averaging interval = 1 minute
Value range:
0...10
Initial value:
10