7.4.3.5 Set Box ID from Scan Header Command
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
0x000B
0x03
0x0005
Used
Ignored
0x00
Bytes 8-11
The four character ASCII
string representing the
logical box number
Byte 12
Null terminator (0x00)
The Vishay StrainSmart application uses this field to save a box number to the header file for the selected
cards. The parameter must be null terminated. If your application involves multiple scanners you may
find it useful to track the box number. Default: 0001
7.4.3.6 Set Project Name to Scan Header File Command
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
varies
0x03
0x0006
Used
Ignored
0x00
Bytes 8-N
The ASCII string
representing the StrainSmart
project which generated this
header file.
The maximum string length
is 48 bytes (including the
null terminator).
Used by Vishay’s StrainSmart application software.
7.4.3.7 Set Scan Session Descriptor to Scan Header File Command
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
varies
0x03
0x0007
Used
Ignored
0x00
Bytes 8-N
The ASCII string
representing the StrainSmart
scan session descriptor
associated with this header
file.
The maximum string length
is 48 bytes(including the null
terminator).
Used by Vishay’s StrainSmart application software.