114
Multimeter Command Reference
SAMPle
The
SAMPle
command subsystem operates with the TRIGger command subsystem.
The
SAMPle
subsystem designates the number of readings (count) made for each
trigger signal received.
Subsystem Syntax
SAMPle
:COUNt <
number
>|MIN|MAX
:COUNt? [MIN|MAX]
:COUNt
SAMPle:COUNt <
number
>|MIN|MAX
designates the number of readings per
trigger.
Parameters
Comments
•
MINimum
sets 1 reading per trigger.
MAXimum
sets 50,000 readings per
trigger.
•
If
MAX
or
50,000
is specified for
number
, the command executes without
error. When an
INIT
is executed requiring readings to be stored in internal
memory, an “Insufficient memory” error is generated to show that the number
of readings exceeds the memory available. However, you can execute
READ?
which returns the readings to the output buffer and does not use internal
memory.
•
A number >50,000 returns Error -222, “Data out of range”.
•
CONFigure
and
MEASure
set the sample count to 1.
•
*RST Condition:
SAMP:COUN 1
Example
Set the Sample Count
CONF:VOLT:DC
Function: DC voltage.
TRIG:SOUR EXT
Trigger source is external BNC on multimeter
front panel.
SAMP:COUN 10
Specify 10 readings per trigger.
READ?
Place multimeter in wait-for-trigger state;
make measurement when external trigger is
received; send readings to output buffer.
enter statement
Enter readings into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
number
>
numeric
1 through 50,000|MIN|MAX
none
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...