88
Multimeter Command Reference
:CURRent[:DC]
CONFigure:CURRent[:DC] [<
range
>|MIN|MAX|DEF|AUTO
[,<
resolution
>|MIN|MAX|DEF]]
selects the DC current function and allows you to
specify the measurement range and resolution.
Parameters
Comments
•
To select a standard measurement range, specify
range
as the input signal's
maximum expected current. The multimeter then selects the correct range to
accept that input.
•
The
AUTO
option for the
range
parameter enables autorange and will not
accept a
resolution
parameter but will default the integration time to 10 PLC.
•
The
DEFault
option for the
range
parameter will also enable autorange.
The
DEF
option for the
resolution
parameter defaults the integration time to
10 PLC.
•
The
MIN
and
MAX
parameters select the minimum or maximum values for
range
and
resolution
:
For
range
:
MIN
= 10 mA;
MAX
= 3A
For
resolution
: See Table 3-1 on page 70 for valid resolution choices for
each range.
•
To select autorange, specify
AUTO
or
DEF
for
range
or do not specify a value
for the range and resolution parameters (see next bullet comment). In the
autorange mode, the multimeter samples the input signal before each
measurement and selects the appropriate range.
•
To specify the
MIN
or
MAX
resolution while autoranging, you must specify the
AUTO
or
DEF
parameter for range and specify
MIN
or
MAX
e.g.,
CONF:CURR:DC DEF,MIN
or
CONF:CURR:DC DEF,MAX
or
CONF:CURR AUTO,MIN
or
CONF:CURR AUTO,MAX
(you cannot omit the
range
parameter
DEF
or
AUTO
). This prevents the
MIN
or
MAX
resolution
from being interpreted as a range setting.
Example
Making
DC
Current Measurements
CONF:CURR 3,MAX
Function: dc current; range selected: 3A;
MAX resolution: 0.3mA.
SAMP:COUN 3
Take 3 readings; trigger source is IMMediate
by default.
READ?
Place multimeter in wait-for-trigger state and
make measurements; send readings to output
buffer.
enter statement
Enter readings into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
10mA|100mA|1 A|3 A|
MIN|MAX|DEF|AUTO
A
<
resolution
>
numeric
resolution|
MIN|MAX|DEF
A
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...