110
Multimeter Command Reference
[:VOLTage[:DC]]:RATio?
MEASure[:VOLTage[:DC]]:RATio? [<
range
>|MIN|MAX|DEF|AUTO
[,<
resolution
>|MIN|MAX|DEF]]
configures the multimeter for dc:dc ratio
measurements with the specified range and resolution. For ratio measurements, the
specified range applies to the signal connected to the HI and LO input terminals.
Autoranging is automatically selected for reference voltage measurements on the
“Sense” HI and LO terminals with a maximum voltage of 10V.
Parameters
Comments
•
To select a standard measurement range, specify
range
as the input signal's
maximum expected voltage. The multimeter then selects the correct range to
accept the input.
•
The
AUTO
or
DEFault
option for the
range
parameter enables autorange.
The
DEF
option for the
resolution
parameter defaults the integration time to
10 PLC.
•
The
MIN
and
MAX
parameters select the minimum or maximum values for
range
and
resolution
:
For
range
:
MIN
= 100mV;
MAX
= 300V.
For
resolution
:
MIN
selects the best resolution (the smallest value) for the
selected range.
MAX
selects the worst resolution (the largest value) for the
selected range.
•
To select autorange, specify
DEFault
for
range
or do not specify a value for the
range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
•
To specify a
MIN
or
MAX
resolution while autoranging, you must specify
MEAS:VOLT:DC:RAT? DEF
(you cannot omit the
range
parameter). This
prevents the
MIN
or
MAX
resolution from being interpreted as a range setting
and the resulting command becomes
MEAS:VOLT:DC:RAT? DEF,MIN
or
MEAS:VOLT:DC:RAT? DEF,MAX
.
•
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making DC Voltage Ratio Measurements
MEAS:VOLT:DC:RAT? 0.825,MAX
Function: DC voltage; range selected: 1V;
MAX resolution: 100
V.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
100mV|1V|10V|100V|300V|
MIN|MAX|DEF|AUTO
volts
<
resolution
>
numeric
resolution|
MIN|MAX|DEF
volts
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...