141
HP-IB Programming
OLA Specific Commands
command
:SENSe:POWer:ATIMe?
syntax
:SENSe[1|2]:POWer:ATIMe?
description
This query returns the current setting of the time interval used for
averaging power measurements. It can be used any time.
parameter
1|2
The channel parameter 1 or 2 may be present, but is ignored.
response
The average time is returned as a number in exponential format:
2E-2, 2E-1, or 1. The unit is seconds. The unit not included in the
response.
related commands
:SENSe:POWer:ATIMe
example
OUTPUT 724;":SENS:POW:ATIM?"
ENTER 724;A$
command
:SENSe:POWer:CALCulate:MODe
syntax
:SENSe[1|2]:POWer:CALCulate:MODe<wsp><mode>
description
This command sets the measurement and display mode for all
polarization dependent measurements, as can be done manually
with the PRESET key.
parameters
1|2
The channel parameter 1 or 2 may be present, but is ignored.
Mode
REFResh|AVERage|0|1. You can enter either the mnemonic
or the corresponding number:
REFResh
0
Refresh
AVERage
1
Average
related commands
:SENSe:POWer:CALC:MODE?
examples
The following examples yield the same instrument setting:
OUTPUT 724;":SENS:POW:CALC:MODE AVER"
OUTPUT 724;":SENS:POW:CALC:MODE 1"