142
HP-IB Programming
OLA Specific Commands
command
:SENSe:POWer:CALC:MODe?
syntax
:SENSe[1|2]:POWer:CALC:MODe?
description
This query returns the current setting of the measurement and
display mode for the polarization dependent applications. It can be
used any time.
parameter
1|2
The channel parameter 1 or 2 may be present, but is ignored.
response
A
boolean
value:
0
Refresh
1
Average
related commands
:SENSe:POWer:CALC:MODE
example
OUTPUT 724;":SENS:POW:CALC:MOD?"
ENTER 724;A$
command
:SENSe:POWer:HEAD?
syntax
:SENSe[1|2]:POWer:HEAD?
description
This query returns information about the optical head connected to
the specified channel. It can be used any time.
parameter
1|2
1 Reads head A
2 Reads head B
Default is 1.
response
An
integer
number:
0
There is no head attached to the specified channel
1
There is a head attached to the specified channel
2
A low PDL head is attached to the specified channel
3
A low PDL head with temperature control is attached to the
specified channel
example
OUTPUT 724;":SENS2:POW:HEAD?
ENTER 724;A$