Section 7. Installation
238
•
When testing and troubleshooting fast measurements, the following
Status
table registers may provide useful information:
o
SkippedScan
(p. 550)
o
MeasureTime
o
ProcessTime
(p. 547)
o
MaxProcTime
(p. 544)
o
BuffDepth
(p. 537)
o
MaxBuffDepth
(p. 544)
•
When the number of
Scan()/NextScan
BufferOptions
is exceeded, a
skipped scan occurs, which means a measurement was missed.
•
Bursts have a duty cycle less than 100%. Assuming no other
measurement instructions are present in the program, each burst occurs at
the beginning of the
Scan()
Interval
. During the rest of the scan, the
CR800 catches up on overhead tasks and processes data stored in buffers.
•
If you wish to account for the time needed in the
Scan()/NextScan
Interval
, consider the following two points:
o
Status
table
MeasureTime
field reports the measurement time
that occupies the
Scan()/NextScan
Interval
.
MeasureTime
includes
time needed to make measurements inside and outside
SubScan()
/
NextSubScan
.
o
NextScan
needs 100 µs to run
•
One
Scan()
/
NextScan
buffer holds the raw measurements made in one
main scan, inside and outside the sub-scan.
For example, one execution of the following code sequence stores 30000
measurements in one buffer:
'Scan(Interval, Units, BufferOption, Count)
Scan
(40,Sec,3,0)
SubScan
(2,mSec,10000)
VoltSe
(Measurement(),3,mV5000,[U6]1,False,150,250,1.0,0)
NextSubScan
NextScan
•
You can dwell burst more than one channel with the same program by
adding a voltage measurement instruction for each channel to be
measured. Channels will be measured in series.
•
The following points apply to cluster burst measurements:
o
Measure smaller clusters for faster speeds.
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......