Abbreviated commands:
Commands can be abbreviated with letters shown in boldface. Letters not printed in boldface can be left out, for example: “
OU
TPUT ON” = “OU ON”
As a rule, letters can be entered in upper or lower case.
Stringing commands together:
Several commands is a single data string must be separated by semicolons “;”, for example: “USET 12; ISET 8.5; OUTPUT ON”
Numeric parameter formats:
m,
n:
Whole number (integer)
v:
Whole number, fixed or floating decimal point number with or without exponent, for example:
“12.5”, “0012.5”, “1.25E1”, “+1.25 e+01”
Function-Specific Commands and Device Settings
trG txt
Function selection for the trigger input at the analog interface
X
unchanged
T_M
ODE
txt
txt:
OFF
(input disabled)
X
X
OUT
(input deactivates or activates the output (OUTPUT OFF / ON))
RCL
(input triggers step-by-step memory recall)
SEQ
(input starts or stops sequential memory recall (SEQUENCE GO / STOP))
LLO
(input enables or disables front panel controls (LOCAL LOCKED))
MIN
(input sets Min-Max storage function (MINMAX OFF / RST or ON))
TS
ET
v
Set memory location-specific default time for the SEQUENCE function
X
X
X unchanged
v:
0; 0.01; ... , 99.99 (delay time in seconds)
X
X
UI_
txt
Min-Max storage for measured U and I values
X
X
OFF
MI
NMAX
txt
txt:
OFF, ON, RST (min-max value storage off / on / reset)
X
X
UL
IM
v
Voltage setting limit (soft-limit)
X
X
U
nominal
v:
15.0
(example: 15.0 V)
X
X
US
ET
v
voltage setpoint
X
X
X 000.000
v:
12.5
(example: 12.5 V)
X
X
W
AIT
v
Additional waiting time
X
X
v:
0.001;0.002 ... 9,999 (waiting time in seconds)
General Commands and Settings
*CLS
Clear status command
X
X
no effect
DCL, SDC
DEVICE CLEAR: Clear input and output buffers (all settings and register contents remain unchanged)
X
X
*
DDT
txt
DEFINE DEVICE TRIGGER: Definition of function to be executed after receiving the
*TRG command or IEC bus signal GET (group execute trigger)
X
X
DDT memory
cleared
txt:
USET 12 / ISET 8.5 / OUT ON / IOUT? (example)
txt
≤
80 characters, use “/” instead of “;” as a delineator within
txt
!
ERAE
n
Device dependent event register A enable command
X
X
no effect
ERBE
n
Device dependent event register B enable command
X
X
no effect
*ESE
n
Standard event enable command
X
X
no effect
IFC
INTERFACE CLEAR: Rest and initialize the IEC bus interface
(all settings and register contents remain unchanged)
X
X
*OPC
Operation complete command
X
X
no effect
*PRE
n
Parallel poll enable register command
X
X
no effect
*PSC
n
Power-on status clear command
X
X
no effect
rcl
n
RECALL: Recall and activate a device setting which has been stored with the *SAV command
X
*
RCL
n
n:
1 ; 2 ; ... , 10 (memory location number)
X
X
11 ; 12 ; ... , 255 (memory location number, only available with SEQUENCE function set to ON)
*
RST
RESET: Reset device to default settings
X
X
X
*SRE
n
Service request enable command
X
X
no effect
SRQ
Service request
X
no effect
Sto n
SAVE: Save momentary device settings to battery-backed memory
X
Memory contents
unchanged
*
SAV
n
n:
0
*SAV 0 clears the content of memory locations
n1
to
n2
by using the address parameters of the
sequence function defined by START_STOP
n1
,
n2
X
X
1 ; 2 ; ... , 10 (memory location number)
11 ; 12 ; ... . 255 (location number for SEQUENCE function, only available with SEQUENCE function set to ON)
*
TRG
TRIGGER: Execute command(s) specified by means of *DDT
X
X
*WAI
Wait to continue command
X
X
no effect
Setting saved with *SAV n:
Expla-
nation
in
chap.
on
page
Manual setting or via interface
n=1
...
10
n=11
...
255
Setting Command
Significance / Effect
Manu
al
IEEE 488
RS 232C
Default setting
after RESET
*RST: