©Isothermal Technology
Page 51 of 61
914 millisKanner manual Iss. 1.00
Isothermal Technology,millisKanner,13-P525,1.00
8.2.6.4
8.2.6.4
8.2.6.4
8.2.6.4
Command: *RST
Command: *RST
Command: *RST
Command: *RST
Format: *RST
Performs a reset (equivalent to a power-on reset) of the millisKanner.
8.2.6.5
8.2.6.5
8.2.6.5
8.2.6.5
Command: [ROUTe:]CLOSe
Command: [ROUTe:]CLOSe
Command: [ROUTe:]CLOSe
Command: [ROUTe:]CLOSe#
#
#
#
Format: [ROUTe:]CLOSe<input#>
{POSitive|NEGative}
Selects (connects input to output) the specified input (on the millisKanner to which the
RS232 connection is made). The polarity of the sense terminals is defined by the
{POSitive|NEGative}
parameter. The millisKanner returns a <CR> (carriage return) to
acknowledge that the channel has been selected and is ready to use.
Example: to select the first millisKanner input with normal (positive) polarity, use:
CLOSE0 POS
8.2.6.6
8.2.6.6
8.2.6.6
8.2.6.6
Command: [ROUTe:]CLOSe?
Command: [ROUTe:]CLOSe?
Command: [ROUTe:]CLOSe?
Command: [ROUTe:]CLOSe?
Format: [ROUTe:]CLOSe?
Reports the connection status (which channels are selected and their settings) on the
millisKanner to which the RS232 connection is made.
Example: a milliKanner with channel 3 selected for normal polarity (positive) will respond
to
CLOS?
with “3,positive”. If no inputs are selected, the response is “0,open”
8.2.6.7
8.2.6.7
8.2.6.7
8.2.6.7
Command: [ROUTe:]OPEN[:ALL]
Command: [ROUTe:]OPEN[:ALL]
Command: [ROUTe:]OPEN[:ALL]
Command: [ROUTe:]OPEN[:ALL]
Format: [ROUTe:]OPEN[:ALL]