Reports the temperature of the reference junction (on the millisKanner to which the
RS232 connection is made)
.
Example: a millisKanner with a reference junction temperature of 20.123°C will respond
to
MEAS:RJC?
with “20.123”.
8.2.6.17
8.2.6.17
8.2.6.17
8.2.6.17
Command: CALibrate:
Command: CALibrate:
Command: CALibrate:
Command: CALibrate:RJC:
RJC:
RJC:
RJC:OFFSet
OFFSet
OFFSet
OFFSet
Format: CALibrate:RJC:OFFSet
Sets the value of the offset applied to the RJC sensor (on the millisKanner to which the
RS232 connection is made). The value is in °C and has a range of ±10°C. This command
will only work if the calibration has first been unlocked using the
CAL:UNL
command (see
section 8.2.6.20).
Example: to set an offset that will be added to the RJC temperature sensor on the
millisKanner to 0.321°C (password still set to default value) use:
CAL:UNL 1234
followed
by
CAL:RJC:OFFS 0.321
8.2.6.18
8.2.6.18
8.2.6.18
8.2.6.18
Command: CALibrate:
Command: CALibrate:
Command: CALibrate:
Command: CALibrate:RJC:
RJC:
RJC:
RJC:OFFSet?
OFFSet?
OFFSet?
OFFSet?
Format: CALibrate:RJC:OFFSet?
Reports the offset added to measurements made by the reference junction sensor (on the
millisKanner to which the RS232 connection is made).
Example: a milliK that has been calibrated to have a 0.321°C offset added to reference
junction sensor measurements will respond to
CAL:RJC:OFFS?
with “0.321”.
8.2.6.19
8.2.6.19
8.2.6.19
8.2.6.19
Command: CALibrate:PASSword
Command: CALibrate:PASSword
Command: CALibrate:PASSword
Command: CALibrate:PASSword
Format: CALibrate:PASSword
<old password>,<password>,<password>