Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
tsplink.trigger[N]
tsplink.trigger[N].assert
Replace N with the number of the synchronization line: 1-3.
Function
Asserts a trigger on one of the synchronization lines.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
tsplink.trigger[n].assert()
n
Trigger line number.
Remarks
The set pulse width determines how long the trigger is asserted.
Details
See
Also see
Example
Asserts trigger on I/O line 2:
tsplink.trigger[2].assert()
tsplink.trigger[N].clear
Replace N with the number of the synchronization line: 1-3.
Function
Clears a trigger event on a synchronization line.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
tsplink.trigger[n].clear()
n
The trigger line number.
Remarks
• The Trigger event detector recalls if a trigger event has been detected since the last
tsplink.trigger[n].wait
call.
• This function clears a trigger event detector, discards the previous history of the trigger line, and
clears the
tsplink.trigger[n].overrun
attribute.
Details
See
Also see
Example
-- Clears trigger event on synchronization line 2:
tsplink.trigger[2].clear()
tsplink.trigger[N].EVENT_ID
Replace N with the number of the synchronization line: 1-3.
Attribute
Identifies the number used for the trigger events used by the tsplink trigger line when it detects an
input trigger.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
event_id = tsplink.trigger[n].EVENT_ID
event_id
The trigger event number.
n
The trigger line.
Remarks
• Set the stimulus of any trigger event detector to the value of this constant to have it respond to
trigger events from this line.