Lake Shore Model 321 Autotuning Temperature Controller User’s Manual
Remote Operation
4-15
Control Process Commands
(Continued)
RAMP
Enable/Disable Ramp Function.
Input:
RAMP 0
or
RAMP 1
Returned:
Nothing
Remarks:
RAMP 0 disables the ramping function while RAMP 1 enables ramping.
RAMP?
Ramping Enable/Disable Status Query.
Input:
RAMP?
Returned:
0
or
1
Remarks:
Returns a 0 if the ramping function is disabled or a 1 if the ramping function is
enabled.
RAMPR
Set Ramp Rate in Kelvin per Minute.
Input:
RAMPR XX.X
Returned:
Nothing
Remarks:
XX.X is the ramp rate in Kelvin per minute between 0 and 99.9.
Example:
RAMP 10[term]
instructs the Model 321 to make the ramp rate equal to 10
K/Min.
RAMPR?
Ramp
Rate
Query.
Input:
RAMPR?
Returned:
XX.X
Remarks:
Returns the current value of the ramp rate.
RAMPS?
Ramping Status Query.
Input:
RAMPS?
Returned:
0
or
1
Remarks:
Returns a 1 if the instrument is in the process of ramping or a 0 if not ramping.