Bytes 0
63
Name
Type of data
Function
Byte 3
Bit 0: error in the application
Bit 1: filling is active
Bit 2: system ready
Bit 3: filling stopped
Bit 4: coarse flow
Bit 5: fine flow
Bit 6: calming time
Bit 7: tolerance alarm
BOOL
Bit 0:
Error bit was set.
Bit 1:
The filling phase is active.
Bit 2:
The system is ready for filling.
Bit 3:
Filling was stopped.
Bit 4:
System is filling: coarse flow bit is set.
Bit 5:
System is filling: fine flow bit is set.
Bit 6:
The bit is set according to the set calming
time.
Bit 7:
alarm if tolerance is above/below set point
Byte 4
7
Gross
DINT/REAL
Weight
Byte 8
11
Net
DINT/REAL
Weight
Byte 12
15
Tare
DINT/REAL
Weight
Byte 16
Material ID
BYTE
Current material ID
Byte 17
Reserved
Byte 18
35
Material name
STRING_18
Current material name
Byte 36
39
Set point
DINT/REAL
Current set point
Byte 40
43
Preset
DINT/REAL
Current set point
Byte 44
47
Material overshoot (OVS)
DINT/REAL
Current set point
Byte 48
51
Calming time
DINT
Current set point
Byte 52
55
Tolerance above set point
DINT/REAL
Current set point
Byte 56
59
Tolerance below set point
DINT/REAL
Current set point
Byte 60
62
Reserved
Byte 63
"Last_Error" error
The last error number is retained until it is
reset (for possible errors see Chap-
ter
).
Transmitter in field housing PR 5230
12 Fieldbus interface
Minebea Intec
EN-276