COMMAND OVERVIEW
17/11/2016 Version No. 002
127
MEASure2:VOLTage:AC?
Returns the AC voltage measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:VOLT:AC?
>+0.10020E-01
Returns the AC voltage measurement as 0.010020V.
MEASure2:CURRent:DC?
Returns the DC current measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:CURR:DC?
>+0.00856E-02
Returns the DC current measurement as 0.0856 mA.
MEASure2:CURRent:AC?
Returns the AC current measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:CURR:AC?
> +0.01254E-02
Returns the AC current measurement as 0.1254mA.