Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
53
Table 52: OUTPut[n]:SCALe
•
< Percentage > := <NRf> value
The value of n indicates the channel number.
Returns
<NRf>
Example
OUTPut1: SCALe 70
Sets the amplitude scale for the channel 1 to 70%
OUTPut1: SCALe?
Might return 70
Command
OUTPut[n]:SERIESIMPedance
Description
This command and query sets or returns the Output Impedance of the
analog channel “n”.
Group
Output
Syntax
OUTPut[n]:SERIESIMPedance {50Ohm|LOW}
OUTPut[n]:SERIESIMPedance?
Related
Commands
None
Arguments
•
50Ohm: the output impedance is set to 50 Ohm
•
LOW: the output impedance is set to Low impedance (5 Ohm)
The value of n indicates the channel number.
Returns
<NRf>
Example
OUTPut1:SERIESIMPedance 50Ohm
Sets the output impedance for the channel 1 to 50 Ohm
OUTPut1:SERIESIMPedance?
Might return LOW that means the channel 1 output impedance is set to
Low value.