Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
96
Description
This command sets and queries the power-on status flag that controls
the automatic power-on execution of SRER and ESER.
When *PSC is true, SRER and ESER are set to 0 at power-on. When
*PSC is false, the current values in the SRER and ESER are preserved
in nonvolatile memory when power is shut off and are restored at
power-on.
Group
Status
Syntax
*PSC {0|1}
*PSC?
Related Commands
Arguments
<NR1>=0
Sets the power-on status clear flag to false, disables the power-on
clear, and allows the instrument to
possibly assert SRQ after power-on.
<NR1>≠0
Sets the power-on status clear flag true; sending *PSC 1 therefore
enables the power-on status clear
and prevents any SRQ assertion after power-on
Returns
A single <NR1> value.
Example
PSC 0
Sets the power-on status clear flag to false.
Table 124: *PSC