Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
100
Command
SYSTem:VERSion? (Query Only)
Description
This command returns the SCPI version number to which the
command conforms.
Group
System
Syntax
SYSTem:VERSion?
Related Commands
Arguments
Returns
A single <NR2> value.
<NR2> ::= YYYY.V where YYYY is the year version and V is revision
number for that year.
Example
SYSTem:VERSion? Might return 1999.0
Table 132: SYSTem:VERSion?
2.22
Trigger Group Commands
Command
ABORt (No Query Form)
Description
This command stops waveform playout when the Run Mode is set to
Gated.
This is equivalent to release the Force Trigger button on the front
panel when the instrument is in gated mode
Group
Trigger
Syntax
ABORt
Related Commands
AWGControl:RMODe, *TRG
Example
ABORt
It will stop the waveform playout on all channels with their Run
Mode set to Gated.
Table 133: ABORt
Command
TRIGger[:SEQuence]:SOURce
Description
This command sets or returns the instrument trigger source.
Group
Trigger
Syntax
TRIGger[:SEQuence]:SOURce {TIMer|EXTernal|MANual}
TRIGger[:SEQuence]:SOURce?
Related Commands
None
Arguments
<source>::= {TIMer|EXTernal|MANual}
TIMer: the trigger is sent at regular intervals.
EXTernal: the trigger come from the external BNC connector.