120
Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Memory Subsystem (:MEMory)
Key Entry
MTONE
Remarks
Refer to
“File Name Variables” on page 13
for information on the file name syntax.
:CATalog:PTRain?
Supported
All with Option 320 and either UNU or UNW
:MEMory:CATalog:PTRain?
This command lists all files of the pulse train files stored in the non- volatile storage.
:CATalog:SEQ?
Supported
N5172B with Option 653 or 655, N5182B with Option 656 or 657
:MEMory:CATalog:SEQ?
This query outputs a list of the arbitrary waveform sequence files. The return data will be in the
following form:
<mem used>,<mem free>,"<file listing>"
The signal generator will return the two memory usage parameters and as many file listings as there
are files in the directory list. Each file listing parameter will be in the following form:
"<file name>,<file type>,<file size>"
Example Output
1818624,519962624,"SEQ1_TEST,SEQ,206","SEQ_TEST,SEQ,169"
Key Entry
SEQ
Remarks
Refer to
“File Name Variables” on page 13
for information on the file name syntax.
:CATalog:SHAPe
Supported
N5172B with Option 653 or 655, N5182B with Option 656 or 657
:MEMory:CATalog:SHAPe?
This command outputs a list of the burst shape files. The return data will be in the following form:
<mem used>,<mem free>,"<file listing>"
The signal generator will return the two memory usage parameters and as many file listings as there
are files in the directory list. Each file listing parameter will be in the following form:
"<file name,file type,file size>"
Key Entry
SHAPE
Remarks
Refer to
“File Name Variables” on page 13
for information on the file name syntax.
:CATalog:STATe?
Supported
All Models