Agilent X-Series Signal Generators SCPI Command Reference
179
System Commands
System Subsystem (:SYSTem)
:PRESet:ALL
Supported
All
:SYSTem:PRESet:ALL
This command sets all states of the signal generator back to their factory default settings, including
states that are not normally affected by signal generator power–on, preset, or
*RST
.
For a comparison of the SCPI preset commands, refer to
Table 3- 1, “Preset SCPI Commands
Overview,” on page 178
.
:PRESet:LANGuage
Supported
All
:SYSTem:PRESet:LANGuage
"SCPI"|"COMP"|"8648"|"E4428C"|"E4438C"|"E8257D"|"E8267D"|"E8663B"|"E8247C"|"E8257C"|"
E8267C"|"N5181A"|"N5182A"|"E442XB"|"E443XB"|"E8241A"|"E8244A"|"E8251A"|"E8254A"|"SMU2
00A"|"SMATE200A"|"SMJ100A"|"SMIQ"|"SML"|"SMV"|"SMR"|"SMF100A"|"MG3691B"|"MG3692B"|"MG
3693B"|"MG3694B"|"3410"|"8360"|"8372"|"83732"|"83752"|"8340"|"8644"|"8662"|"8663"|"86
64"|"8665"
:SYSTem:PRESet:LANGuage?
This command sets the remote language that is available when the signal generator is preset.
SCPI
This choice provides compatibility for SCPI commands.
COMP
This choice provides compatibility for the 8656B, 8657A/B signal generator which
is supported only through the GPIB interface.
8648
This choice provides compatibility for the 8648A/B/C/D signal generator which is
supported only through a GPIB interface.
:SYSTem
:PON:TYPE
PRESet|LAST|USER
:SYSTem:PON:TYPE?
Sets the power on state (PON) to be the same
as the front- panel green Preset hardkey, or
the last state, or to the user state.
:SYSTem
:PRESet
Performs the same preset as currently set for
the front- panel green Preset hardkey.
:SYSTem
:PRESet:USER:SAVE
Saves the current instrument state as the user
preset state.
:SYSTem
:PRESet:ALL
Sets the instrument to the same default
conditions performed by sequentially
inputting:
:SYSTem:PRESet
+
:SYSTem:PERSistent:PRESet
:SYSTem
:PRESet:USER
Executes a user preset.
Table 3-1
Preset SCPI Commands Overview (Continued)
Command
Description
Remarks