Agilent X-Series Signal Generators SCPI Command Reference
23
Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
Remarks
The setting enabled by this command is not affected by signal generator power–on,
preset, or
*RST
.
:PMETer:COMMunicate:LAN:DEVice
Supported
All Models
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:DEVice <deviceName>
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:DEVice?
This command enters a VXI–11 device name for a power meter that is being controlled by the signal
generator for user flatness calibration. If connecting directly to the power meter, enter the name as
specified on your power meter documentation. If connecting through a LAN–GPIB gateway, enter the
SICL address of the power meter.
Key Entry
PM VXI–11 Device Name
Remarks
The setting enabled by this command is not affected by signal generator power–on,
preset, or
*RST
.
:PMETer:COMMunicate:LAN:IP
Supported
All Models
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:IP <ipAddress>
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:IP?
This command sets the internet protocol (IP) address for a power meter that is controlled by the
signal generator for user flatness calibration. If connecting to a GPIB power meter through a
LAN–GPIB gateway, this command sets the IP address of the gateway.
Key Entry
Power Meter IP Address
Remarks
The setting enabled by this command is not affected by signal generator power–on,
preset, or
*RST
.
Ensure that the power meter IP address is different from the signal generator
address.
:PMETer:COMMunicate:LAN:PORT
Supported
All Models
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:PORT <portNumber>
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:PORT?
This command sets the IP port number on the power meter that is controlled by the signal generator
for users flatness calibration.
Key Entry
Power Meter IP Port
5025
Standard mode. The command enables standard mode for simple programming.
5024
Telnet mode. The command enables the telnet SCPI service for programming.