Agilent X-Series Signal Generators SCPI Command Reference
System Commands
System Subsystem (:SYSTem)
Remarks
After setting a new preset language and presetting the instrument, some defaults
may change to the preset values of the instrument indicated by the remote
language.
:PRESet:PERSistent
Supported
All
:SYSTem:PRESet:PERSistent
This command sets the states that are not affected by signal generator power–on, preset, or
*RST
to
their factory default settings.
Key Entry
Restore System Settings to Default Values
Remarks
For a list of the persistent instrument factory default values refer to the
Programming Guide
.
For a comparison of the SCPI preset commands, refer to
Table 3- 1, “Preset SCPI
Commands Overview,” on page 178
.
:PRESet:TYPE
Supported
All
:SYSTem:PRESet:TYPE NORMal|USER
This command defines the Preset hardkey as either factory preset or as the user preset saved in
memory.
NORMal
This choice uses the factory–defined defaults when
Preset
is pressed.
COMP
This choice uses the user–defined preset saved in the instrument when
Preset
is
pressed. Refer to
:PRESet:USER
and
:PRESet:USER:SAVE
commands.
Remarks
This command will return an error, if the USER parameter is sent without a user
preset saved in the instrument.
:PRESet:USER
Supported
All
:SYSTem:PRESet:USER
This command presets the signal generator to the user’s saved state.
Remarks
This command presets the signal generator to the saved user–defined state.
For a comparison of the SCPI preset commands, refer to
Table 3- 1, “Preset SCPI
Commands Overview,” on page 178
.
Key Entry
SCPI
SMJ100A
8648A/B/C/D
E8257D, E8267D, E8663B
SML
3410 Series
8656B, 8657A/B
E8241A, E8244A, E8251A, E8264A
SMV
E4428C,E4438C
SMU200A, SMATE200A
8662A
SMIQ
E442xB, E443xB
E8247C, E8257C, E8267C
8663A
Key Entry
Preset
Key Entry
Execute User Preset