26
Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Digital Modulation Subsystem—N5172B/82B ([:SOURce])
Digital Modulation Subsystem—N5172B/82B ([:SOURce])
:BURSt:STATe
Supported
N5172B/82B
[:SOURce]:BURSt:STATe ON|OFF|1|0
[:SOURce]:BURSt:STATe?
This command enables or disables the burst envelope function.
*RST
0
Key Entry
Burst Envelope Off On
:DM:CORRection:OPTimization
Supported
N5172B/82B
[:SOURce]:DM:CORRection:OPTimization RFOut|EXTernal
[:SOURce]:DM:CORRection:OPTimization?
This command enables the internal optimized path to accommodate I/Q signals.
EXT
This choice applies correction terms to provide a calibrated signal at the IQ
output. When the I/Q Output is selected, the RF signals at the RF Output are
uncalibrated.
RFO
This choice applies correction terms to provide a calibrated signal at the RF
output. When the RF Output is selected, the I/Q signals at the I/Q Output are
uncalibrated.
*RST
RFO
Key Entry
Correction Optimized Path
:DM:EXTernal:INPut:ATTen
Supported
N5172B/82B
[:SOURce]:DM:EXTernal:INPut:ATTen <val><unit>
[:SOURce]:DM:EXTernal:INPut:ATTen?
This command sets the attenuation level for the external I/Q signals being modulated through the
signal generator RF path. The variable <val> is expressed in decibels (dB).
Example
:DM:EXT:INP:ATT 10
The preceding example sets the modulator attenuator to 10 dB.
*RST
+2.00000
Range
0
to
40 dB
Key Entry
Modulator Atten Manual Auto