Agilent X-Series Signal Generators SCPI Command Reference
201
Analog Modulation Commands
Amplitude Modulation Subsystem–Option UNT ([:SOURce])
:AM[1]|2:INTernal:SWEep:FUNCtion:SHAPe:RAMP
Supported
All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:SHAPe:RAMP POSitive|NEGative
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:SHAPe:RAMP?
This command sets the ramp direction for the swept function generator when
:AM[1]|2:INTernal:SWEep:FUNCtion:SHAPe
is set to RAMP.
*RST
POS
:AM[1]|2:INTernal:SWEep:RATE[1]|2
Supported
All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:RATE[1]|2 <val><units>
[:SOURce]:AM[1]|2:INTernal:SWEep:RATE[1]|2?
This command sets the sweep rate for the AM swept–sine waveform.
The sweep rate function is only available for internal source 1.
*RST
+4.00002
Range
0.5 Hz to 100 kHz
Key Entry
AM Sweep Rate
:AM[1]|2:INTernal:SWEep:SHAPe
Supported
All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:SHAPe SAWTooth|TRIangle
[:SOURce]:AM[1]|2:INTernal:SWEep:SHAPe?
This command selects the sweep shape.
SAWTooth
A sawtooth sweep shape only sweeps from start freq to stop freq.
TRIangle
A triangle sweep sweeps from start freq to stop freq and back to start freq. With
a triangle shape sweep you can choose a different sweep rate for start freq to
stop freq and another sweep rate for stop freq to start freq.
*RST
SAWTooth
:AM[1]|2:INTernal:SWEep:TIME:COUPled
Supported
All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:TIME:COUPled ON|OFF|1|0
[:SOURce]:AM[1]|2:INTernal:SWEep:TIME:COUPled?
Key Entry
Pos Ramp
Neg Ramp
Key Entry
AM Sweep Shape