13-4
SCPI Signal Oriented Commands
Model 2750 Multimeter/Switch System User’s Manual
Query
CONFigure?
Query the selected function.
Description
<clist> included
— When the <clist> parameter is included with
CONFigure command, the specified channel(s) for the scanlist assumes
the *RST default settings for the specified function. Range can also be
set for the channel(s) by including the <rang> parameter. If the
resolution parameter (<res>) is included, it will be ignored.
The present measurement function and the trigger model settings are not
affected when the CONFigure command is sent with the <clist>
parameter.
<clist> not included
— When the <clist> parameter is not included, the
CONFigure command configures the instrument for subsequent mea-
surements on the specified function. Range and resolution can also be
set for the specified function.
This command places the instrument in a “one-shot” measurement
mode. You can then use the READ? command to trigger a measurement
and acquire a reading (see
).
When this command is sent without the <clist> parameter, the Model
2750 will be configured as follows:
•
The function specified by this command is selected. If specified,
range and/or resolution are also set.
•
All controls related to the selected function are defaulted to the
*RST values.
•
Continuous initiation is disabled (INITiate:CONTinuous OFF).
•
The control source of the Trigger Model is set to Immediate.
•
The count values of the Trigger Model are set to one.
•
The delay of the Trigger Model is set to zero.
•
The Model 2750 is placed in the idle state.
•
All math calculations are disabled.
•
Buffer operation is disabled. A storage operation presently in pro-
cess will be aborted.
•
Autozero is enabled.
Programming examples:
Programming example
#1 — The following command configures
scanlist channels 101 through 105 for 4-wire resistance measurements
on the 1M
Ω
range.
CONF:FRES 1e6, (@101:105)
Programming example #2
— The following command selects the
DCV function, 10V range, 3
H
digit resolution, and performs the “no
<clist>” CONFigure operations:
CONF:VOLT 10, 3.5
2750-900-01.book Page 4 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...