Model 6485 Picoammeter Instruction Manual
DDC Emulation Commands
D-5
SRQ
(cont.)
M25
Busy, Reading Done or Reading Overflow
*ESE 1
STATus:MEASurement[:EVENt]:ENABle 192
*SRE <
value
>
value
= (current *SRE value) |= (0x20|0x01)
(B4, B3,
B0)
H
M26
Busy or Reading Done or Data Store Full
*ESE 1
STATus:MEASurement[:EVENt]:ENABle 576
*SRE <
value
>
value
= (current *SRE value) |= (0x20|0x01)
(B4, B3,
B1)
NEW, H
M27
Busy or Reading Done or Data Store Full or Reading
Overflow
*ESE 1
STATus:MEASurement[:EVENt]:ENABle 704
*SRE <
value
>
value
= (current *SRE value) |= (0x20|0x01)
(B4, B3,
B1, B0)
NEW, H
M32
Clears SRQ Error Mask
*ESE 0
(this disables all events in the Standard Event
Register)
*SRE <
value
>
value
= (current *SRE value) &= (~ESB)
(None)
X
M33
IDDCO
(B5, B0), I
M34
IDDC
(B5, B1), I
M35
IDDC or IDDCO
(B5, B1,
B0), I
M36
Not in Remote
(B5, B2), I
M37
Not in Remote or IDDCO
(B5, B2,
B0), I
M38
Not in Remote or IDDC
(B5, B2,
B1), I
M39
Not in Remote, IDDC or IDDCO
*ESE 48
Enables the SCPI command and Execution
error events in the standard Event Register
*SRE <
value
>
value
= (current *SRE value) |= (ESB). This Enables an
SRQ on vents enabled in the Standard Event Register.
(B5, B2,
B1, B0), I
Table D-1 (continued)
Device dependent command summary
Mode
Command
Description
Equivalent SCPI commands
Note
Default
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......